Advantages

  • Optimum protection at all times through microprocessor-controlled real-time automatic speed regulation.
  • Audio and visual alarms with error messages for security.
  • Noise level lower than 54 dBA for user comfort.
  • Easy maintenance from the front.
  • Compliant with NF EN 12469:2000.

BSC TYPE II: UCS-Basic TechGen

Principle

The UCS Basic has one self-regulating fan to ensure a constant recycled air / extracted air ratio and two H14 HEPA filters (one for air circulation in the workspace, the other for extraction to the exterior).

70% of air is recycled through the main filter and 30% is exhausted through the extraction filter.

The filtration system maintains a protective barrier at a minimum velocity of 0.4 m/s and a Class ISO 3 workspace according to NF ISO 44201.

Construction

Epoxy painted sheet metal exterior.
Stainless steel workspace.
Stainless steel work surface.
Spill tray.
2 H14 HEPA filters (efficiency > 99.995% MPPS). .

Dimensions

ModelEffective
dimensions (cm)
Overall
dimensions (cm)
 LPHLPH
UCS-Basic 73 38 58 83 65 93

The ucs-basic techgen is microprocessor-controlled

Control panel with digital keyboard and 2-line, 20 character LCD screen.

Operating parameter display:
Speeds, airflow, residual lifetime of filters, etc.

Audio and visual alarms:
Abnormal speed, fan or sensor malfunction, HEPA filter clogging, sash position, etc.

Secure standby mode.

USC-Basic TechGen microprocesseur

1 IP44 electric socket.

Lighting in excess of 1200 lux.

Filter test inlet.
To ensure the integrity of HEPA filters.

Perforated single-piece work surface.

Fluids :

  • tap (gas, vacuum, water, N2 on request)
  • additional socket
  • RJ45, USB, RS 232 ports

Hanging bar.
To hang bags.

UV lamp with timer and alarm on closure panel.

Aluminium closure panel.

Armrest.

Support stands:

  • Epoxy painted steel.
  • Stainless steel.
  • With casters.
  • Electrically adjustable.