Avantages
- Optimum protection at all times through microprocessor-controlled real-time automatic speed regulation.
- Audio and visual alarms with error messages for security.
- Double-sealed on three sides for maximum protection.
- Sash tiltable to 130° on hydraulic cylinders for cleaning.
- Easy maintenance from the front.
- Compliant with NF EN 12 469:2000 and has the NF label as per NF-095.
Principle
The SCS TechGen has two self-regulating fans to ensure a constant recycled air / extracted air ratio and two H14 HEPA filters (one for air circulated into the workspace, the other for extraction to the exterior).
70% of air is recycled through the main filter and 30% is exhausted through the extraction filter.
The filtration system maintains a protective barrier at a minimum velocity of 0.4 m/s and a Class ISO 3 workspace according to NF ISO 44201.
Construction
Epoxy painted sheet metal exterior.
Stainless steel workspace.
Stainless steel work surface.
Spill tray.
Tiltable sash for cleaning.
Double-seal negative pressure plenum and air intake on the three sides.
Available in one- and two-fan versions with automatic speed control.
H14 HEPA filters (efficiency > 99.995% MPPS).
Dimensions
Model | Effective dimensions (cm) | Overall dimensions (cm) | ||||
---|---|---|---|---|---|---|
L | P | H | L | P* | H | |
*Reducible to 75cm to pass through doorways. | ||||||
BSC EN 3 | 89 | 58 | 66 | 105 | 79 | 147 |
BSC EN 4 | 119 | 58 | 66 | 135 | 79 | 147 |
BSC EN 5 | 149 | 58 | 66 | 166 | 79 | 147 |
BSC EN 6 | 180 | 58 | 66 | 196 | 79 | 147 |
THE BSC-EN TECHGEN IS MICROPROCESSOR-CONTROLLED
Control panel with digital keyboard and 2-line, 20 character LCD screen.
Operating parameter display:
Speeds, airflow, residual lifetime of filters, etc.
Audio and visual alarms:
Abnormal speed, fan or sensor malfunction, HEPA filter clogging, sash position, etc.
Secure standby mode.
1 IP44 electric socket.
Lighting in excess of 1200 lux.
Protection grid.
To prevent fan failure.
Filter test inlet.
To ensure the integrity of HEPA filters.
Autoclavable work surface.
Solid or perforated, segmented or one-piece, at the user’s choice at no additional cost.
Fluids :
- tap (gas, vacuum, water, N2 on request)
- additional socket
- RJ45, USB, RS 232 ports
Hanging bar.
To hang bags.
Antivibration platform.
Accurate to 0.1 mg.
Additional exhaust filter (two-fan version).
- active carbon filter
- HEPA filter
Connection kit.
Volt-free contact.
For on/off relay.
UV lamp with timer and alarm.
Armrest.
Support stands:
- Epoxy painted steel.
- Stainless steel.
- With casters.
- Electrically adjustable.