Advantages

  • Optimum protection at all times through microprocessor-controlled real-time automatic speed regulation.
  • Audio and visual alarms with error messages for security.
  • Optimum ergonomic design with electrically operated sliding sloped sash and glass side panels.
  • Sash tiltable to 130° on hydraulic cylinders for cleaning.
  • Easy maintenance from the front.
  • Compliant with NF EN 12469:2000.

BSC TYPE II B: TCS TechGen

Principle

The TCS TechGen is a Type II B BSC. It has two self-regulating fans to ensure a constant recycled air / extracted air ratio and three filters (H14 HEPA filter or carbon filter for recovery under the work surface, H14 HEPA filter for air circulated into the workspace and H14 HEPA filter for venting to the exterior).

100% of air is recovered under the work surface, then 70% is recycled through the main filter and 30% is discharged through the extraction filter.

The filtration system maintains a protective barrier at a minimum velocity of 0.4 m/s and a Class ISO 3 workspace according to NF ISO 44201.

Construction

Epoxy painted sheet metal exterior.
Stainless steel workspace.
Stainless steel work surface.
Spill tray.
Electrically operated sliding sloped sash, tiltable for cleaning.
Safety-glass side panels.
Available in two-fan version with automatic speed control.
H14 HEPA filters (efficiency > 99.995% MPPS).

Dimensions

ModelEffective
dimensions (cm)
Overall
dimensions (cm)
 LPHLP*H
*Reducible to 79cm to pass through doorways.
TCS 3 89 58 74 105 86 196
TCS 4 119 58 74 135 86 196
TCS 5 149 58 74 166 86 196
TCS 6 180 58 74 196 86 196

The tcs techgen is microprocessor-controlled

Control panel with digital keyboard and 2-line, 20 character LCD screen.

Operating parameter display:
Speeds, airflow, residual lifetime of filters, etc.

Audio and visual alarms:
Abnormal speed, fan or sensor malfunction, HEPA filter clogging, sash position, etc.

Secure standby mode.

TCS TechGen microprocesseur

1 IP44 electric socket.

Pre-drilled holes in side panels.
(3 per side)

Lighting in excess of 1200 lux.

Protection grid.
To prevent fan failure.

Filter test inlet.
To ensure the integrity of HEPA filters.

Autoclavable work surface.
Solid or perforated, segmented or one-piece, at the user’s choice at no additional cost.

Fluids:

  • tap (gas, vacuum, water, N2 on request)
  • additional socket
  • RJ45, USB, RS 232 ports

Hanging bar.
To hang bags.

Antivibration platform.
Accurate to 0.1 mg.

Additional exhaust filter (two-fan version).

  • active carbon filter
  • HEPA filter

Active carbon filter to replace the HEPA filter under the work surface.

Bag In/Bag Out system for the filter under the work surface.

Connection kit.

Volt-free contact.
For on/off relay.

UV lamp with timer and alarm.

Armrest.

Support stands:

  • Epoxy painted steel.
  • Stainless steel.
  • Electrically adjustable.